HIOKI ST5541 Leak Current HiTester
Leak current measurement, an essential part of electrical safety (For Standard- and Regulation-compliant)
LEAK CURRENT HiTESTER ST5541: A low-cost solution that complies with standards governing general-use electrical devices. Uninterrupted polarity switching function dramatically reduces cycle time. Support for rated currents up to 20 A gives the instrument more than adequate capability for testing products designed to comply with new standards. Touch panel features simple, interactive operation. Communications functionality and external I/O support allow automatic testing on production lines.
Note: For applications involving leak current measurement of medical-use electrical devices, use the ST5540.
Basic specifications (Accuracy guaranteed for 1 year)
Measurement methods : Measurement of voltage drop across body simulated resistance points
Calculation and display of current values True rms measurement
Measurement unit floats relative to instrument ground.
Measurement modes : Leak current measurement, voltage measurement, safety conductor current measurement
Standard compliance : Electrical Appliances and Materials Safety Act
Measurement of touch current and protective conductor current: IEC 60990 (1999)
Electrical equipment for measurement, control, and laboratory use: IEC 61010-1 (2001)
Information technology equipment: IEC 60950-1 (2005)
Household and similar appliances: IEC 60335-1 (2001) + A1:2004 + A2:2006
Audio, video and similar electronic apparatus: IEC 60065 (2001) + A1:2005
Personnel Protection Systems for EV: UL-2231-1 (2002), UL-2231-2 (2002)
UL: UL-1492 (1996)
Electrical equipment for measurement, control, and laboratory use; current measurement circuits in damp conditions: IEC 61010-1 (2001)
Leak current measurement : Ground leak current, 3 types of contact current, free current measurement, 3 types of enclosure leak current
Measurement current : DC, AC (true rms, 15 Hz to 1 MHz), AC+DC (true rms, 15 Hz to 1 MHz), AC peak (15 Hz to 1 MHz)
Measurement ranges : DC / AC / AC+DC mode:50 µA/500 µA/5 mA/50 mA
AC peak mode:500 µA/1 mA/10 mA/75 mA
Measurement accuracy(current measurement) : DC measurement: ±2.0% rdg. ±6 dgt. (typ.)
AC / AC+DC measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 100 kHz, typ.)
AC peak measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 10 kHz, typ.)
Interfaces : External I/O, USB 1.1 (communications), RS-232C
Functionality : Automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc.
HiTESTER power supply : 100/120/220/240 V AC (specify at time of order), 50/60 Hz, 30 VA rated power
Target device power supply input : 100 to 240 V AC, 50/60 Hz Rated current input from terminal block: 20 A
Target device power supply output : Output from terminal block: 20 A Output from outlet: 15 A
Dimensions, mass : 320 mm (12.60 in) W × 110 mm (4.33 in) H × 253 mm (9.96 in) D, 4.5 kg (158.7 oz)
Supplied accessories : TEST LEAD L2200 (Red ×1, Black ×1) ×1 set, ENCLOSURE PROBE 9195 ×1, Instruction manual ×1, Users manual ×1, Power cord ×3, Spare fuse for measurement line ×1