HIOKI RM3542-01 Resistance HiTester
Measure in as little as 0.9 ms, Optimized for Automated Systems
High speed and accuracy maximize productivity in automated systems. Multiple checking functions ensure proper contact for reliable measurements. Low-power resistance mode measures chip inductors and EMC suppression components. Supports sample inspections during the manufacturing process.
Test fixtures are not supplied with the unit. Please select an optional test fixture when ordering
Basic specifications (Accuracy guaranteed for 1 year) (Warm-up time 30 minutes or more)
Resistance range : [at Low Power OFF] 100 mO range (Max. 120.0000 mO, 0.1µO resolution)) to 100 MO range (Max. 120.0000 MO, 100 O resolution), 10 steps
[at Low Power ON] 1000 mO range (Max. 1200.000 mO, 1 µO resolution) to 1000 O range (Max. 1200.000 O, 1 mO resolution), 4 steps
Display : Monochrome graphic LCD 240 × 64 dot, white LED backlight
Measurement accuracy : [with SLOW mode, at 100 mO range] ±0.015 % rdg. ±0.002 % f.s.
[with SLOW mode, at 1000 O range] ±0.006 % rdg. ±0.001 % f.s. (the best case)
Testing current : [at 100 mO range] 100 mA DC to [at 100 MO range] 100 nA DC
Open-terminal voltage : 20 V DC max.
Sampling rate : FAST, MEDIUM, SLOW, 3 settings
Measurement times : [at 100 O /1000 O ranges, with Low Power OFF] FAST: 0.9 ms, MED: 3.6 ms, SLOW: 17 ms (minimum time)
Integration time : 0.1 ms to 100.0 ms, or 1 to 5 PLC at 50 Hz, 1 to 6 PLC at 60 Hz
Note: PLC = one power line cycle (mains wave-form period)
Other functions : Comparator (compare setting value with measurement value), Delay (set to allow for mechanical delay of trigger input and probing, or set to allow for measurement object response), OVC (offset voltage compensation), Measurement fault detection, Probe short-circuit detection, Improve contact, Auto-memory, Statistical calculations, Settings monitor (when using two instruments, a difference in settings causes warning notification), Retry, Trigger function. etc,.
Interface : RS-232C, Printer, GP-IB (Model RM3542-01)
External I/O : Trigger, Hold input, Comparator output, Settings monitor terminal
Power supply, consumption : 100 to 240 V AC, 50/60 Hz, 30 VA max.
Dimensions, mass : 260 mm (10.24 in) W × 88 mm (3.46 in) H × 300 mm (11.81 in) D, 2.9 kg (102.3 oz)
Supplied accessories : Power cord × 1, EXT I/O male connector × 1, Operation manual × 1, Operation guide × 1